Eldico Blog

Quality breakthrough: R1<10% on Triphenylbenzene

Written by Dr. Eric Hovestreydt | May 3, 2022 6:47:03 AM

Triphenylbenzene experiments on ELDICO ED-1 brought our continuous R&D to a significant milestone: R values as low as 9% were achieved with a standard kinematical refinement from a single run.

Triphenylbenzene (C24H18) is a compound of interest found at the intersection of chemistry and physics applications. A versatile photoluminescent chemo-sensor with environmental and biological importance, it can assist with its polymeric forms to efficiently reduce low-concentration CO2 .

During one of the numerous experiments on ELDICO’s ED-1 and as part of our continuous fine-tuning of the device, we were happy to attain low R values, which signal that complementary quality to X-ray diffraction experiments is possible in nano-sized compounds while skipping the crystal growth stage.

Experimental setting

Data processing and refinement within established Apex (® Bruker) software package.  Standard kinematical refinement without taking dynamic scattering into account. Data of one single phi-scan were used.

Crystal Parameters:

Chemical formula

C24H18

Crystal system

Orthorhombic P

Space group

Pna21

a, b, c (Å)

7.60(4), 19.68(11), 11.25(6)

Volume (Å3)

1681(16)

Z

4

F(000)

648


Refinement Details:

R1 (%)

9.70

wR2 (%)

28.05

GoF

1.15

Rint (%)

5.6

Resolution (Å)

0.99

Completeness (%)

63

Independent reflections

1069

Δρmin, Δρmax

-0.09, 0.09

 

‘For me this particular experiment was thrilling as I was able to see comparable data to the industry standard attained with X-rays - now on electrons. Not to forget that this is a significant step forward for a technique that can measure very, very small crystals without the hassle of growing large enough crystals’, said Gunther Steinfeld, experienced crystallographer and ELDICO’s founder.